· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Mega Soft Universal Plus Patient Return Electrode | 0847 |
| All lots. | Mega2000 Patient Return Electrode | 0800 |
| All lots. | Mega2000 Soft Patient Pad Return Electrode | 0835 0830 |
| All lots. | Mega Soft Universal Patient Return Electrode | 0846 0840 0845 |
| All lots. | Mega Soft Dual Universal Plus Patient Return Electrode | 0848 |
Megadyne Medical Products, Inc. (“Megadyne”) has received reports of patient burns identified after surgical procedures in which Mega Soft pads were used. Megadyne is aware of 63 complaints of serious patient burns globally since April 2018. Megadyne has conducted a thorough investigation, and to this point have not identified any design or manufacturing defects, nor determined definitive root cause. However, Megadyne has determined that in some instances the Mega Soft pad Instructions for Use (IFU) were not being properly followed.
Recall start date: June 6, 2023