· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Contact the manufacturer. | Eclipse Mobility Scooters | S148 S846 S940GTX S840GT S746 |
There is a possibility that a small portion of one seat pin batch under the seat base of the affected mobility scooter(s) was not forged in the right conditions. This may result in insufficient strength and may cause a breakage. In the event of an accident where the seat pin breaks, the occupant may fall and cause an injury.
Recall start date: May 18, 2021