Health Products

8mm, Tip-Up Fenestrated Grasper, IS4000

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: 8mm, Tip-Up Fenestrated Grasper, IS40
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.8mm, Tip-Up Fenestrated Grasper, IS4000470347-12

Issue

The tip-up fenestrated grasper has two versions on the approved product list (APL): 470347-11 and 470347-12. PN 470347-11 uses machined grips, while PN 470347-12 uses grips that have been manufactured via metal-injection molding (MIM). There have been complaints reported for a broken grip issue against the MIM grips. Grip breakage is observed on both upper and/or lower grips.

Recall start date: June 8, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime