Health Products

Philips Azurion and Allura Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Philips Azurion and Allura Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog numbe
Not applicable.Allura Xper FD20 (Product Of AlluraClarity Family)722028
Not applicable.Allura Xper FD10001443
Not applicable.Allura Xper FD20 Or Table (Part Of AlluraClarity Family)722035
Not applicable.Model or catalog numbe722 224
722 079
Not applicable.Allura Xper FD10 (Product Of AlluraClarity Family)722026

Issue

Due to the forces applied during the movement of the C-arc of the Azurion and Allura systems, the bolts supporting the flexmove carriage may become loose and/or break, and cracks may appear in the flexmove carriage.

Recall start date : June 12, 2023