· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog numbe |
|---|---|---|
| Not applicable. | Allura Xper FD20 (Product Of AlluraClarity Family) | 722028 |
| Not applicable. | Allura Xper FD10 | 001443 |
| Not applicable. | Allura Xper FD20 Or Table (Part Of AlluraClarity Family) | 722035 |
| Not applicable. | Model or catalog numbe | 722 224 722 079 |
| Not applicable. | Allura Xper FD10 (Product Of AlluraClarity Family) | 722026 |
Due to the forces applied during the movement of the C-arc of the Azurion and Allura systems, the bolts supporting the flexmove carriage may become loose and/or break, and cracks may appear in the flexmove carriage.
Recall start date : June 12, 2023