Health Products

BD Synapsys

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: BD Synapsys
Issue: Medical devices - Performance
What to do: Review recall notice, share/post within facility, complete/return reply form, scheddule remote servicing.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Version 2.1 and higherBD Synapsys444150

Issue

BD recently identified, through a customer complaint investigation, that there is a risk of mis-association of patient demographic data for BD Synapsys™ users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (skip, final discard, mark as read, save) to process patient results rapidly.

Recall start date : June 8, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime