Health Products

Simplexa Direct Amplification Disc Kit

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Microbiology
Product: Simplexa Direct Amplification Disc Kit
Issue: Medical devices - Performance
What to do: Review notice & return reply form, discard remaining inventory, and retest any sample with an error code.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected prodcuts
More than 10 numbers, contact manufacturer.MOL1455Simplexa Direct Amplification Disc Kit

Issue

A recent internal investigation identified a specific set of Simplexa Direct Amplification Discs (Mol1455) lots produced with a specific lot of raw material are at potential risk of leaking when the disc is re-used multiple times with the assays Simplexa HSV 1&2 Direct Mol2150, Simplexa Bordetella Direct (Mol2750), and Simplexa VZV Direct (Mol3650). A total of 13 affected lots of Mol1455 were shipped and distributed in Canada. A total 15 customers running these assays have the affected disc lots.

Recall start date: June 26, 2023