· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All Lots. | System One A-Series Heated Humidifier | 1111552 1077107 1111555 |
| All lots. | System One Heated Humidifier | IN6T INX3T IN6TFLG IN6H INX3H IN6HFLG |
| All Lots. | BiPAP A40 Pro, CA W/ Heated Tube Humidifier, CA | CAX3100T12 |
| All lots. | BiPAP A30 System - Ventilator & System One Aseries Heated Humidifier | 1111148 |
| All lots. | BiPAP A40 EFL | INX3T |
| All lots. | BiPAP A40 Pro, CA | CAX3100S12 |
| All lots. | BiPAP A40, Canada, Core Package | 1111174 |
There is currently a non-conformance regarding the disinfection method recommended in the instructions manual for devices in the field, as the Keredusy disinfection method was not adequately validated. Long term effects of this method on the device components are unknown. Full validation requires biological evaluation after cleaning cycles per ISO 10993-1 and testing on expected worse-case clinical use per ISO 18562-1. the maximum number of Keredusy cycles was also not identified within the user manuals. ISO 17664-1 does allow for the manufacturer's IFU to direct the processor to refer to the disinfecting agent manufacturer's IFU with reference to concentration, temperature, and contact time, but these parameters were not defined nor documented within the IFU. The IFU only directed the user to Philips customer service.
Recall Start Date: June 30, 2023