· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 710223E00008 710223E00017 710223E00010 710223E00016 710223E00007 710223E00014 710223E00011 710223E00013 710223E00012 710223E00009 | Statclave G4 Chamber Autoclaves | G4-622301 |
An error in production resulted in missing gaskets in some units on float sensors in the reservoirs. The units were not leaking when they left the facility, however a leak will develop over time with a potential to cause an electrical short.
Recall Start Date: June 16, 2023