Health Products

Statclave G4 Chamber Autoclaves

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Statclave G4 Chamber Autoclaves
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
710223E00008
710223E00017
710223E00010
710223E00016
710223E00007
710223E00014
710223E00011
710223E00013
710223E00012
710223E00009
Statclave G4 Chamber AutoclavesG4-622301

Issue

An error in production resulted in missing gaskets in some units on float sensors in the reservoirs. The units were not leaking when they left the facility, however a leak will develop over time with a potential to cause an electrical short.

Recall Start Date: June 16, 2023