Health ProductsType I recall

Evis Exera Video System and OES - Bronchoscopes

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ear, nose and throat, Health products - Medical devices - Gastroenterology and urology
Product: Evis Exera Video System and OES - Bronchoscopes
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.OES BronchofiberscopesBF-P60
BF-1T60
BF-MP60
All lots.Evis Exera III Video System - BronchovideoscopeBF-Q190
BF-1TH190
BF-H190
BFXT190
All lots.Evis Exera II Video System - BronchovideoscopesBF-1TQ180
BF-P180
BF-Q180-AC
BF-Q180
BF-1T180
All lots.Evis Exera Video Sytem - BronchovideoscopeBF-XT160

Issue

Olympus has received complaints of endobronchial combustion during therapeutic procedures using lasers or argon plasma coagulation with the Olympus bronchoscope. If endobronchial combustion occurs, patients may suffer internal burn in their airway or lungs, respiratory insufficiency, apnea, loss of consciousness, hospitalization or its prolongation, ICU care, or death.

Three (3) adverse event complaints with endobronchial combustion during laser or argon plasma coagulation procedures have occurred, of which one (1) complaint resulted in patient death. In an effort to maximize patient safety and mitigate any potential risk to patient health, Olympus is notifying users of these complaints and the providing the following recommendations related in combination with laser therapy equipment: - Only Nd:YAG laser or 810 nm diode lasers may be used with Olympus laser compatible bronchoscopes. Olympus has not evaluated any other lasers for compatibility with the indicated bronchoscope models - Do not perform laser cauterization while supplying oxygen. This may result in combustion during cauterization. This is included in the warnings in the operation manual on laser cauterization with Olympus bronchoscopes. - Never emit laser radiation before confirming that an appropriate distance between the target and the endoscope's distal end with the tip of the laser probe is in the correct position in the endoscopic image. This is essential to avoid patient injury (burns, bleeding, & perforation) or damage to the device.

Recall start date: June 30, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime