· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | HF-Resection Electrode | WA22740S WA22755S WA22644S WA22655S WA22640S WA22744S |
| All lots. | HF-Resection Electrode | WA22543S WA22305S WA22541S WA22566S WA22542S WA22544S WA22540S |
| All lots. | HF-Resection Electrode | WA22655C WA22355A WA22355C |
| All lots. | HF-Resection Electrode | WA22521C WA22623C WA22621C WA22523C |
| All lots. | HF-Resection Electrode | WA47738S WA47751S |
| All lots. | HF-Resection Electrode | WA22632D |
| All lots. | HF-Resection Electrode | WA47741S WA47766S |
| All lots. | HF-Resection Electrode | WA22641S WA22666S WA22766S WA22741S |
| All lots. | HF-Resection Electrode | WA22332D WA22603D WA22507D WA22602D WA22306D WA22607D WA22537D WA22503D WA22558C WA22606D |
| All lots. | HF-Resection Electrode | WA47760S |
| All lots. | HF-Resection Electrode | WA47506S WA47540S WA47507S WA47560S WA47555S WA47505S WA47551S WA47566S |
| All lots. | HF-Resection Electrode | WA22603S WA22742S WA22606S WA22702S WA22732S WA22643S WA22602S WA22637S WA22703S WA22705S WA22743S WA22706S WA22739S WA22607S WA22639S WA22737S WA22632S WA22642S WA22707S WA22605S |
| All lots. | HF-Resection Electrode | WA47705S WA47737S WA47732S WA47707S WA47739S WA47742S WA47743S WA47706S |
| All lots. | HF-Resection Electrode | WA22723S WA22621S WA22721S WA22623S |
| All lots. | HF-Resection Electrode | WA22651S WA22751S WA22638S WA22738S |
| All lots. | HF-Resection Electrode | WA22557C WA22657C |
| All lots. | HF-Resection Electrode | WA22760S WA22657S |
| All lots. | HF-Resection Electrode | WA22538C WA22651C |
| All lots. | HF-Resection Electrode | WA47721S WA47723S |
| All lots. | HF-Resection Electrode | WA47755S WA47740S WA47744S |
| All lots. | HF-Resection Electrode | A22221C A22224C A22255C A22201D A22259C WA22305D A22205D A22206C A22223C A22231C A22266C WA22351C A22251C A44728C WA22038C WA22302D A22206D A22207C WA22331D A22201C A22205C A22211C A22222C A22253C A22257C A22257D WA22037C A22202D A22250C WA22351A WA22539D A22202C A22231D A22258C WA22039C WA22301D A22203C |
Olympus has initiated this recall after receiving complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved.
Recall start date: June 23, 2023.