· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Not applicable. | Azurion 7 M12 | 722 078 |
| Not applicable. | Allura Xper FD10 | 001443 |
| Not applicable. | Allura Xper FD20/15 | 722058 |
| Not applicable. | Allura Xper FD20 (Product Of Alluraclarity Family) | 722028 |
| Not applicable. | Azurion 7 M20 | 722 224 722 079 |
| Not applicable. | Allura Xper FD20/20 (Product Of Alluraclarity Family) | 722038 |
| Not applicable. | Allura Xper FD20 Or Table (Product Of Alluraclarity Family) | 722035 |
| Not applicable. | Allura Xper FD10 (Product Of Alluraclarity Family) | 722026 |
Philips has become aware of several situations that can result in a loss of availability of the wireless foot switch during a procedure. Identified issues that may result in the wireless foot switch not being available are: • Loss of bluetooth connection due to interferences from other radio equipment • Battery not fully charged • Battery not holding its charge • Damage in the charger, cable and/or connector of the wireless foot switch
Recall start date: July 4, 2023