| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Constellation Vision System Fragmentation Pak | 8065-7509-58 |
| More than 10 numbers, contact manufacturer. | Centurion Vision System - Active Irrigation FMS Pack, Infusion Sleeves | 8065752918 |
| More than 10 numbers, contact manufacturer. | Centurion Vision System - Active Irrigation FMS Pack, With Intrepid Ultra Sleeves | 8065752201 8065752200 |
| More than 10 numbers, contact manufacturer. | Turbosonics Tip | 8065-7900-22 8065-7508-53 8065-7408-37 8065752065 8065752066 8065-7900-20 |
| More than 10 numbers, contact manufacturer. | Custom-Pak Ophthalmic Surgery Procedure Pack | 00000-00 TO 99999-99 |
| More than 10 numbers, contact manufacturer. | Intrepid Plus FMS Pack, Ultra Infusion Sleeve, Mini-Flared ABS Tip | 8065752087 |
Alcon Canada Inc. will be initiating a voluntary medical device recall (advisory) on specific catalog numbers of Alcon Phaco tips & packs containing those Phaco tips. There is a potential for the Phaco tip wrench to generate plastic particles if the Phaco tip is over-tightened to the handpiece. The purpose of the advisory notice is to provide customer guidance when tightening the Phaco tip to mitigate the risk of generating plastic particles that have potential to enter the eye during surgery.
Recall start date: June 30, 2023