Health Products

Olympus Pulmonary Endoscopes

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology, Health products - Medical devices - Ear, nose and throat, Health products - Medical devices - Gastroenterology and urology, Health products - Medical devices - General and plastic surgery
Product: Olympus Pulmonary Endoscopes
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
All lots.BF-1TQ180
BF-P180
BF-Q180-AC
BF-Q180
BF-1T180
Evis Exera III Video System - Bronchovideoscope
All lots.BF-MP160F
BF-XP160F
Evis Exera Video System - Bronchofibervideoscopes
All lots.MAF-GMAirway Mobilescope- General
All lots.MAF-TMAirway Mobilescopetherapeautic
All lots.MAF-DM2
MAF-GM2
Airway Mobilescope
All lots.BF-UC190FEvis Eus Ultrasound Bronchofibervideoscope
All lots.LF-PTracheal Intubation Fiberscope
All lots.BF-3C160
BF-XT160
Evis Exera Video Sytem - Bronchovideoscope
All lots.BF-MP190FEvis Exera III Bronchofibervideoscope
All lots.BF-P60
BF-N20
BF-1T60
BF-XP60
BF-MP60
BF-3C40
Oes Bronchofiberscopes
All lots.LF-DP
LF-P
LF-GP
LF-TP
Tracheal Intubation Fiberscopes
All lots.BF-XT190
BF-P190
BF-1TH190
BF-H190
BF-Q190
BF-XP190
Evis Exera III Video System - Bronchovideoscope
All lots.BF-UC180FEvis Exera II Ultrasonic Bronchofibervideoscope
All lots.MAF-TM2Airway Mobilescope
All lots.LF-VVisera Trachael Intubation Videoscope

Issue

Olympus has received three complaints, including one that was associated with an injury, that upon trying to use an Olympus pulmonary endoscope model with an endotracheal tube, the tip became lodged (entrapped) inside the endotracheal tube connector. As a result of the complaint investigation, it was determined that the scope (bending section) was too large for the endotracheal tube connector. Selection of an endotracheal tube where the size is incompatible with the Olympus airway mobilescope can result in delayed procedures, foreign body obstruction, failure to ventilate the patient, hemorrhage, upper airway tract injury, esophageal perforation, and/or airway mobilescope damage.

In an effort to maximize patient safety and mitigate any potential risk to patient health, Olympus is providing the following information related to using an endotracheal tube: • before inserting the airway mobilescope with an endotracheal tube into the patient, slide the tube along the entire length of the airway mobilescope insertion section to confirm that the insertion section of the airway mobilescope can be inserted smoothly into the endotracheal tube. If it cannot be inserted smoothly, the covering material of the bending section of the airway mobilescope or the external surface of the insertion section may be damaged. When using a lubricant, make the above confirmation before applying the lubricant.

In the event the endotracheal tube cannot slide smoothly over the insertion section of the airway mobilescope, select another size tube, or inspect scope for damage.

Olympus is undertaking this action to provide an IFU addendum which specify the correct maximum scope diameter information.

Recall start date: July 28, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime