Health Products

Aquilex Fluid Control System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Aquilex Fluid Control System
Issue: Medical devices - Performance
What to do: Contact the manufacture if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.AQL-100PBS
AQL-100P
Aquilex Fluid Control System

Issue

WOM has received complaints from customers in the United States that the display of inflow volume on the Aquilex Fluid Control System can reach its limit of 30,000 ml during long procedures. In these cases, the display does not indicate the actual inflow volume. In cases where an unusually high amount of fluid is used for distension of the patient's uterus, the calculation of the inflow and outflow volumes reaches its limit and the pump control unit display will freeze at the maximum value. There is a risk of distension fluid reaching the circulatory system of the patient's soft tissue.

Recall start date: July 17, 2023