Health Products

Liquid Assayed Chemistry Control Premium Plus

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Liquid Assayed Chemistry Control Premium Plus
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
1308UELiquid Assayed Chemistry Control Premium Plus1308UE

Issue

Randox laboratories can confirm that there has been a transcription error in the instructions for use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3, LAE4215, lot 1308UE. For caeruloplasmin, the one and two standard deviation values printed on the sheet are incorrect. The target and range values are correct. The IFU has been updated and is available on www.randox.com. Incorrect IFU should be discarded and the updated IFU should be downloaded.

Recall Start Date: August 3, 2023

Get told about the next health product recall.