Health Products

PDS II (Polydiaxanone) Absorbable Suture

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: PDS II (Polydiaxanone) Absorbable Suture
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.PDS II (Polydiaxanone) Absorbable SuturePDP442H
PDP497G

Issue

Ethicon received seven complaints for one lot of undyed PDS™ II sutures regarding low tensile strength. Internal testing on returned product from this lot confirmed that some PDS™ II sutures from this lot did not meet Ethicon's tensile strength requirement. An investigation identified additional lots across the PDS™ Plus Antibacterial and PDS™ II suture families that have the potential to be affected by the same root cause as the lot that received the complaints.

Recall Start Date: August 8, 2023