· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Excor System - Arterial Cannulas | C60G-002M C85G-004 C80G-040M C85G-002M C60G-004M C80G-021M |
| All lots. | Excor System - Connecting Set For Cannulas | A06-009 A09-012 |
| All lots. | Excor System - Apex Cannulas | C14A-040M C18A-020 C22A-004 C27A-001 |
| All lots. | Excor System - Atrial Cannulas | C15V-040M C23V-004 C25V-004 C22V-002M C19V-020M C22V-004 |
The purpose of the urgent medical device correction notification is to advise healthcare professionals that Berlin Heart is conducting a voluntary correction concerning all Excor cannulae due to the potential of partial or complete breaches of an Excor cannula. A significant cannula breach may result in massive blood loss and/or an air embolism which could result in blood loss and/or death to the patient. No product will be removed.
Recall Start Date: August 4, 2023