Health Products

Combolab AltiX

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Combolab AltiX
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
SVS23040008SA
SVS23040003SA
Combolab AltiX2300000-005

Issue

GE Healthcare has become aware that a diode on the power supply of the Prucka 3 Amplifier used with cardiolab/combolab systems could reach elevated temperatures. This could cause the amplifier to fail and consequently power off and become inoperable.

Recall Start Date: August 8, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime