· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| 65236202 65249215 | 00-7711-006-40 00-7711-004-40 | Zimmer M/L Taper Taperfemoral Stem |
Zimmer Biomet is conducting a lot specific medical device recall for two lots of the Zimmer M/L taper hip prosthesis. The outer package labeling and product etch are for a size 6, however, the implant is a size 4, and vice versa.
Recall start date: August 2, 2023