Health Products

Zimmer M/L Taper Taperfemoral Stem

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Zimmer M/L Taper Taperfemoral Stem
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
65236202
65249215
00-7711-006-40
00-7711-004-40
Zimmer M/L Taper Taperfemoral Stem

Issue

Zimmer Biomet is conducting a lot specific medical device recall for two lots of the Zimmer M/L taper hip prosthesis. The outer package labeling and product etch are for a size 6, however, the implant is a size 4, and vice versa.

Recall start date: August 2, 2023

Get told about the next health product recall.
Get health product recall alerts