· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Trexo Home | A0001 |
| More than 10 numbers, contact manufacturer. | Trexo Plus | A0001 |
The following malfunctions and nonconformances were observed during Trexo lab-testing and Trexo design reviews of the device, Resulting in this voluntary device recall and proposed field correction. Electromagnetic (EM) radiation levels above threshold (Ref: IEC 60601-1-11), as observed during lab-testing contracted by Trexo. Water ingress was observed within electronics enclosures after IP22 testing, as observed during lab-testing contracted by Trexo.
Recall start date: July 31, 2023