· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Serial Number 70020 4041 2221 | Pentacam® AXL Wave | 70020 |
| Serial Number 68120 8411 0221 | Myopia Master® | 68120 |
We have found an insufficient anti-reflective coating on an optical component, which can under certain circumstances lead to an additional axis length signal. If the correct axis length signal is weaker, an incorrect axis length value can be displayed.
Recall start date: August 11, 2023