Health Products

Pentacam® AXL Wave and Myopia Master®

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ophthalmology
Product: Pentacam® AXL Wave and Myopia Master®
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Serial Number 70020 4041 2221Pentacam® AXL Wave70020
Serial Number 68120 8411 0221Myopia Master®68120

Issue

We have found an insufficient anti-reflective coating on an optical component, which can under certain circumstances lead to an additional axis length signal. If the correct axis length signal is weaker, an incorrect axis length value can be displayed.