Medfusion Syringe Infusion Pumps

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Medfusion Syringe Infusion Pumps
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Medfusion Syringe Infusion Pumps4000-0101-50
4000-0105-51
4000-0106-01
4000-0100-50
4000-0101-51
All lots.Medfusion Syringe Infusion Pumps3500VX-500
3500-500

Issue

The calibration of the force sensor used to detect occlusions may shift over time. If the force sensor calibration shift is large enough, the pump will display a system failure alarm (including force sensor BGND test, force sensor bridge test, or force sensor test). If the calibration shift is not large enough to trigger a system failure alarm, there may be a slight increase in the threshold to detect an occlusion. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly.

Recall start date: August 8, 2023