· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Vitros 5.1 FS Chemistry System (New & Refurbished) | 6801375 6801890 |
| More than 10 numbers, contact manufacturer. | Vitros 5600 Integrated System (New & Refurbished) | 680 2413 680 2915 |
| More than 10 numbers, contact manufacturer. | Vitros 5.1 FS Chemistry System (New & Refurbished) | 6801375 6801890 |
| More than 10 numbers, contact manufacturer. | Vitros XT 3400 Chemistry System | 684 4458 |
| More than 10 numbers, contact manufacturer. | Vitros 4600 Chemistry System (New & Refurbished) | 6900440 6802445 |
| More than 10 numbers, contact manufacturer. | Vitros 4600 Chemistry System (New & Refurbished) | 6900440 6802445 |
| More than 10 numbers, contact manufacturer. | Vitros 3600 Immunodiagnostics System (New & Refurbished) | 680 2914 680 2783 |
| More than 10 numbers, contact manufacturer. | Vitros XT 7600 Integrated System | 684 4461 |
Certain proboscis/piston assembly (or assemblies) manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an out-of-specification barrel component. This may cause Vitros® Versatips and/or Vitros® microtips to be improperly seated on the barrel component of the assembly, which may result in an increased frequency of condition codes related to the actions of tip pickup, aspirate, dispense, mix, tip-seal, and discard. due to the small Vitros® Microtip and Versatip® dimension variability (within manufacturing tolerances) and the way these tips interact with the assembly, Assemblies manufactured within the affected date range may not perform as intended based on the tips currently in use.
Recall start date: August 10, 2023