Health Products

Essenta Dr Compact Digital Radiography System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Essenta Dr Compact Digital Radiography System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Not applicable.9890 010 88021Essenta Dr Compact Digital Radiography System

Issue

There is a potential cybersecurity issue with Essenta Dr Compact Systems, which could result in unauthorized disclosure or modification of patient data. This issue may occur if a firewall is not in place, the file transfer protocol (ftp) is utilized, and the following sequence of events occurs: • unauthorized personnel gains physical access to customer facility • customer facility does not have a firewall and/or network segregation to isolate medical devices, and • the unauthorized person connects their device to the Essenta Dr Compact System network at the specific time data is transferred (data is not continuously transferred).

Under these conditions unauthorized access or modification of patient data is possible.

Recall start date: August 18 , 2023