Health Products

TC-I Reusable Tip Clip Sensors

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: TC-I Reusable Tip Clip Sensors
Issue: Medical devices - Performance
What to do: Customer to contact Masimo to arrange return and replacement of affected devices.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
Contact manufacturer.Tc-I Reusable Tip Clip Sensors (RD Set TC-I (Oximetry Sensor))4053
Contact manufacturer.Tc-I Reusable Tip Clip Sensors (LNCS Oximetry Sensor)1895
Contact manufacturer.Tc-I Reusable Tip Clip Sensors (M-LNCS Reusable Sensors)2503

Issue

Masimo identified TC-I Reusable Tip Clip Sensors with higher than expected variability in spO2 readings that are outside of the accuracy specification. Masimo's investigation found the TC-I sensors had higher variability with an overall arms of 4.4% compared to the specification of 3.5% arms.

Recall Start Date: June 15, 2023