Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120v
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
C22AKC0070Hugo™ Robotic-Assisted Surgery (RAS) Tower, 120vMRASC0003

Issue

Reports of system error during preoperative calibration self-tests of the Hugo robotic arm cart assembly. After recalibration attempt, the robotic arm cart assembly will be prevented from entering teleoperation even if the calibration is successful. This causes procedural delays in teleoperation.

Recall start date: August 24, 2023