Zimmer Dermatome Blades

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Zimmer Dermatome Blades
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
65630969
65925347
65935738
65709066
65988623
65972711
65599469
65648460
65952858
65952857
00-8800-000-10Zimmer Dermatome Blades

Issue

Zimmer Biomet Canada is initiating a lot specific recall for the Zimmer dermatome blades (00-8800-000-10) in Canada. There have been 38 complaints received related to skin grafts being thin and non-uniform when using the affected blades. The issue would be identified at the time of use and may result in the need for additional harvests to adequately cover the area. The trade-off of incomplete coverage versus having additional grafts is at the discretion of the health care provider based on the condition of the graft, overall patient condition, and severity of the need for the graft.

The investigation identified an assignable cause within the manufacturing process, which was present for a defined amount

Recall start date: September 12, 2023