Health Products

In2Flow Nasal Cannulas

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: In2Flow Nasal Cannulas
Issue: Medical devices - Performance
What to do: Customers provided instructions on how to respond when this becomes a problem.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
1900059952
1900062390
1900059288
1900086945
In2Flow Nasal Cannulas10076606
10076605
10076604

Issue

Failure description: several complaints were received where the plug (the one at the opposite side of the tubing) became detached from the In2Flow nasal cannula, causing leakage due to unintentional disconnection of that part. This may result in degraded high flow support and oxygen delivery, which may lead to patient O2 desaturation.

Recall Start Date: August 31, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime