· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | CombiDiagnost R90 | 709030 |
If the CombiDiagnost R90 System has a damaged wire in the System Power Distribution Unit (SPDU) within the m-cabinet and is connected to the ground, or the system experiences a power surge, there is the potential for a short circuit of the Electromagnetic Interference (EMI) filter or UA/UB board. If this issue occurs, the user may experience one or more of the following: - electrical fire within the main-cabinet (m-cabinet) - see or smell smoke - hear a popping sound coming from the m-cabinet
Recall start date: October 28, 2021