· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | RAPIDPoint® 500 System - Main Unit | 10492730 |
| All lots. | RAPIDPoint® 500e Blood Gas System | 11416751 |
Siemens Healthcare Diagnostics Inc. has determined that two drugs, perhexiline maleate and atomoxetine hydrochloride, may interfere with sodium results that are reported on the RAPIDPoint® 500 and RAPIDPoint® 500e Blood Gas Systems.
Recall start date: October 2, 2023