Astral Ventilators

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Astral Ventilators
Issue: Medical devices - Performance
What to do: Manufacturer provided instructions on how to response. Will implement corrections when available.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Contact the manufacturer.Astral Ventilators27003
Contact the manufacturer.Astral Ventilators27001

Issue

Resmed's Astral ventilator is designed to deliver a Total Power Failure (TPF) audible alarm for at least 2 minutes to alert the carer/clinician when the ventilator experiences a total loss of power that results in a stop of ventilation. The TPF audible alarm is powered by an electrical component known as a supercapacitor. This component degrades over time, which may cause the TPF alarm to sound for less than 2 minutes, or in some cases, to not sound at all.

Recall start date: September 13, 2023