Health ProductsType I recall

Staysafe Disinfection Cap

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: Staysafe Disinfection Cap
Issue: Medical devices - Labelling and packaging, Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
All lots.2845091Staysafe Disinfection Cap

Issue

At the end of August 2022, Fresenius complaint handling unit (Germany) registered a complaint from New Zealand that an infant (no day fill) developed iodine induced hypothyroidism following use of stay safe disinfection cap. Further investigations revealed that the harm of hypothyroidism was not considered in master harms list and was not part of risk analysis for stay safe disinfection cap. Thus, no warning to the user was given. Even though the risk of hypothyroidism and hypersensitivity is known to medical practitioner, related warnings are not provided in the labeling.

Recall start date: September 28, 2023

Get told about the next health product recall.
Get health product recall alerts