Health ProductsType I recall

Cardiohelp-I

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: Cardiohelp-I
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.70104.8012Cardiohelp-I

Issue

During a complaint investigation it was reported that the rotation of the Cardiohelp Emergency Drive hand crank was blocked or impaired. In course of internal investigations, it was found that the safety washer causes this issue. The safety washer was originally introduced to the Cardiohelp Emergency Drive in order to lengthen the shaft of the emergency drive by the thickness of the safety washer reducing the axial pressure on the circlip, safety washer and bearing. (serial # specific)

Recall start date: October 5, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime