· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| 30918Y600 30610Y600 21101Y600 21058Y600 | 6K95-41 6K95-30 | Iron |
| 30227Y600 21118Y600 30302Y600 21046Y600 | 08P3920 | Alinity C Iron Reagent Kit |
Sentinel Ch Spa, the manufacturer of iron for use with Architect and Alinity C Iron Reagent Kit, has identified an issue with the reagents. Some customers have experienced elevated quality control (QC) results as well as elevated patient results. A high recovery of iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of quality control and patient samples results, up to 15% for iron products (Alinity C Iron Reagent ref. 08p3920) to be used on Abbott Alinity instrument and 30% for iron products (iron ref. 6k95-41 and iron ref. 6k95-30) to be used on Abbott Architect instrument.
Recall start date: September 29, 2023