· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| 00802526548406 00802526575167 00802526599002 00802526544101 00802526575228 00802526575105 00802526575136 00802526575143 00802526575112 00802526575211 00802526575129 00802526575181 00802526575204 | A209 | Emblem S-Icd Pulse Generator |
| 00802526584404 00802526590405 00802526590436 00802526581519 00802526584411 00802526590429 | A219 | Emblem Mri S-Icd Pg |
Boston Scientific Corporation (BSC) is initiating a product advisory notification for the Emblem S-ICD pulse generator and Emblem MRI S-ICD pg to inform customers that a software update for the Emblem S-ICD is available to correct the potential for a rare interaction between the Emblem S-ICD and latitude communicator, which may cause S-ICD sensing disablement for a 24-hour interval.
Recall start date: October 2, 2023