| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Laryngeal Injection Needle | 8598B 8598A |
| All lots. | Flexible Injection Needle | 27182A |
| All lots. | Injection Needle | 26175 PD 810506 8598D |
Validation testing of the efficacy of the manual cleaning for the injection needle was performed. The testing showed that the required efficacy level was not achieved. Different scenarios were tested, and all possible manual cleaning steps included, and the residual protein value remained above the acceptance criteria.
Recall start date: October 16, 2023