Evis Exera III Video System - Gastrointestinal Videoscope

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Evis Exera III Video System - Gastrointestinal Videoscope
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lotsEvis Exera III Video System - Gastrointestinal VideoscopeGIF-1TH190

Issue

The GIF-1TH190 reprocessing manual has instructions for either High Level Disinfection (HLD) reprocessing or Ethylene Oxide Gas (EOG) sterilization after endoscope cleaning. As a result of testing, Olympus discovered EOG sterilization failures in which the GIF-1TH190 failed to achieve sterilization following the GIF-1TH190's reprocessing manual. As a result of the subsequent investigation, Olympus determined that changes are required to the endoscope's channel drying steps following cleaning and prior to EOG sterilization in order to ensure effective endoscope sterilization. Accordingly, Olympus has updated the instructions for the endoscope channel's drying process and validated that the new, updated channel drying steps will result in GIF-1TH190 endoscope EOG sterilization. The revised instructions for the "dry the endoscope" in chapter 5.7 of the GIF-1TH190 reprocessing manual are provided in a supplied addendum.

Recall start date: October 11, 2023