· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| Not applicable. | ACHIEVA 1.5T | Achieva 1.5t Mri System-Main Unit |
Philips has identified an issue where a specific component failure in the gradient coil of the affected 1.5T MR systems listed in this letter may act as a heat source with a potential to produce smoke and/or fire. If the component failure occurs, the user may observe the following:
Repeated scan abort error messages due to gradient amplifier fault detection, described in section 4.
Unusual noise from the system in the examination or technical room
Smoke and/or fire from within the system
Recall start date: October 25, 2023
Update (2025-06-04): Philips has identified additional MR systems (Intera 1.5T).