Health Products

Elekta Infinity, Elekta Synergy and Versa Hd

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Elekta Infinity, Elekta Synergy and Versa Hd
Issue: Medical devices - Performance
What to do: Contact ELEKTA_GLOBAL_PMS@ELEKTA.COM if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
No lot numbersXRT1412
XRT1402
Elekta Synergy
No lot numbersXRT1501
XRT1511
Elekta Infitiny
No lot numbersXRT 20319
XRT 20419
Versa Hd

Issue

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iviewgt™ / xvi detector arms. when the detector arm is fully deployed there is no risk of uncontrolled extension.

Recall start date: October 25, 2023