Novum Iq Large Volume Pump Unapproved Adhesive

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Novum Iq Large Volume Pump Unapproved Adhesive
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.40700BAXNovum Iq Large Volume Pump

Issue

Baxter Corporation is issuing an urgent medical device correction for specific Novum Iq LVPs where an unapproved adhesive was used in error during service for the re-installation of the slide clamp assemblies, which can result in in the slide clamp not ejecting properly or the door not opening.

Recall start date: October 26, 2023