Health Products

VITROS Chemistry Products CKMB Slides

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: VITROS Chemistry Products CKMB Slides
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.VITROS Chemistry Products CKMB Slides800 1133
805 8232

Issue

Per the VITROS CKMB slides instructions for use (IFU) "principles of the procedure" section, the spread layer of the CKMB slide contains goat antihuman CK-mm antibodies, which inhibit CK-mm (muscle) activity and ~50% of the CK-mb (heart) activity. The remaining CK activity represents 50% of the total CK-mb isoenzyme activity plus any CK-bb (which is relatively rare). Per the current VITROS CKMB slides IFU "known interferences" section, a total CK activity greater than 1000 u/l may result in falsely elevated CK-mb results. It is limited at 1000 u/l by design, which is due to the level of goat antihuman CK-mm antibodies added during manufacturing. Samples with total CK >1000 u/l should be diluted prior to analysis. Therefore, CK will not typically impact VITROS CKMB results below the CK result of 1000 u/l. Ortho Clinical Diagnostics (Quidelortho) has confirmed the issue and determined that lots manufactured from coating 0261 of VITROS CKMB slides do not adequately inhibit CK-mm up to a total CK >1000 u/l (as stated in the IFU).

Recall Start Date: October 30, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime