Health Products

Stat Profile Prime® Plus Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology, Health products - Medical devices - Chemistry
Product: Stat Profile Prime® Plus Analyzer
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Model or catalog number/Lot or serial numberAffected products
63369Not applicable.Stat Profile Prime® Plus POC Analyzer (Remanufactured)
57400
59508
PP1218410C
PP14210410C
PP1918190C
PP1C18130C
PP12210030C
PP12210080C
PP1420880C
PP15230050C
PP16220090C
PP1918340C
PP1419210C
PP14230260C
PP12230670C
PP18220170C
PP18220090C
PP1219230C
PP1420180C
PP1319060C
PP15230030C
PP18220130C
PP18220180C
PP19181902
PP1918321C
PP1B18340C
Stat Profile Prime® Plus Analyzer
63368PP31220460C
PP33230230C
Stat Profile Prime® Plus POC Analyzer

Issue

Nova Biomedical became aware of a potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new bun/creatinine sensor cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional quality control practices.

Recall start date: September 13, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime