· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 2315316 2315317 | Master Vacuum Pumps | 30.00.30 30.00.35 17-8689 |
When the push-start board was manufactured, a capacitor was incorrectly mounted (polarity). This can lead to overheating if the device is running for min. 60 minutes and then restarted. Delay for restart and treatment.
Recall start date: November 6, 2023