Health ProductsType I recall

Pump, Medfusion, Model 4000

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Pump, Medfusion, Model 4000
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Contact the manufacturer4000-0101-249
4000-0101-50
4000-0105-51
4000-0601-01
4000-0106-00
4000-0100-50
4000-0101-51
4000-0100-249
Pump, Medfusion, Model 4000

Issue

Smiths Medical identified legacy software anomalies which may occur in the Medfusion 4000 syringe infusion pumps. The legacy software anomalies were identified during a historical review of the risk evaluations conducted during the software development process. The historical review determined that some of the risk evaluations conducted in the past did not fully account for all the risks associated with the issues. After re-evaluating the patient risk, Smiths Medical determined that the issues identified were associated with sufficient patient risk that it is necessary to notify users, even if the remediation for an issue was available in previous software versions.

Recall start date: November 7, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime