Health Products

Vitros Chemistry Products System - Eco2 Slides (Carbon Dioxide)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Vitros Chemistry Products System - Eco2 Slides (Carbon Dioxide)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
5732-3598- 4651
5732-3598-4650
826 2396Vitros Chemistry Products System - Eco2 Slides (Carbon Dioxide)

Issue

Ortho Cinical Diagnostics (quidelortho) has determined that some lots from coating 3598 of Vitros Eco2 Slides include carts which may contain consecutive slides that may not function as intended. These affected Vitros Eco2 slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased results. Due to the nature of the abnormality, which can vary across the affected slides, the magnitude of potential bias can vary from slide to slide.

Recall start date: October 26, 2023

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime