Health Products

Synchromed II - Programmable Pump

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Synchromed II - Programmable Pump
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Synchromed II - Programmable Pump8637-20
8637-40

Issue

Medtronic recently identified that if the Synchromed II Pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field. In this case, drug delivery will only resume after performing a post MRI pump interrogation with the clinician programmer which will end telemetry mode.

Recall Start Date: November 21, 2023