· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Power LED 300 Surgical Light | ARDXXXXXXXXXX |
| All lots. | Axcel | ARDXXXXXXXXXX |
| All lots. | Hanaulux HLX3000 | ARDXXXXXXXXXX |
| All lots. | Blueline Series 30/80 | ARDXXXXXXXXXX |
| All lots. | Maquet Equipment | ARDXXXXXXXXXX |
| All lots. | Axcel+ | ARDXXXXXXXXXX |
| All lots. | Lucea 50/100 | ARDXXXXXXXXXX |
| All lots. | Blue Series 30/80 Surgical Light | ARDXXXXXXXXXX |
| All lots. | PrismAlix | ARDXXXXXXXXXX |
| All lots. | Maquet Power LED II | ARDXXXXXXXXXX |
| All lots. | Volista Surgical Light | ARDXXXXXXXXXX |
| All lots. | Power LED Surgical Light | ARDXXXXXXXXXX |
| All lots. | Lucea 10/40 | ARDXXXXXXXXXX |
| All lots. | X'Ten Surgical Light | ARDXXXXXXXXXX |
| All lots. | H Led Surgical Light | ARDXXXXXXXXXX |
Maquet S.A.S. /Getinge has received a customer complaint regarding a fall of a light system in an operating room in Hong Kong (no injuries reported). Tnvestigation into this event demonstrated that the recommendation of maintenance and service were not performed as instructed in the current user and maintenance manuals. This notification communicates retrospectively current maintenance instructions and programs to customers.
Recall start date: November 29, 2023