Health ProductsType I recall

Alaris™ GH Plus Guardrails™ Syringe Pump

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Alaris™ GH Plus Guardrails™ Syringe Pump
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected ProductsModel or catalog number
More than 10 numbers, contact manufacturer.Alaris™ GH Plus Guardrails™ Syringe Pump8002MED01-G

Issue

BD is issuing a notification to inform customers that Cardinal Health branded Monoject™ syringes have not been validated for use with the BD Alaris™ infusion devices. BD has validated Covidien branded Monoject™ syringes for use with the BD Alaris™ infusion devices. Cardinal Health, rebranded its syringes from Covidien Monoject™ to Cardinal Health Monoject™ and retained the original Covidien SKU numbers. BD's investigation determined that there may differences in syringe dimensions associated with these changes that may impact BD Alaris infusion devices.

Recall Start Date: December 7, 2023