· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 22F0950JZX 22K0967JZX 22E1619JZX 22J0164JZX | Cytosponge™ Cell Collection Device | CYTO-201 |
Specific lot numbers of Cytosponge Cell Collection device may be at increased risk of the sponge detaching from the string during removal of the device from the patient.
Recall Start Date: June 2, 2023