· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Scorpio Total Knee - Concentric Dome Patellar Component | 73-2110 73-2910 73-2710 |
| More than 10 numbers, contact manufacturer. | Scorpio Total Knee - Resurfacing Patellar Component | 73-0510 73-0110 73-0710 73-0908 |
| More than 10 numbers, contact manufacturer. | Scorpio Total Knee - Universal Dome Patellar Component | 73-3110 73-3308 73-3710 73-3508 73-3708 73-3910 |
| 33164901 | Duracon Total Knee System - PCA Review - Tibial Insert | 6637-4-231 |
| 42912101 42995601 42874701 42961901 | Scorpio NRG Knee System - Cruciate Retaining Tibial Insert | 82-2-0910 82-2-0908 |
| 38059601 | Trident Acetabular Shell - Conventional Polyethylene | 620-00-28J |
| 42771401 41875401 42771601 42579701 | Duracon Total Knee System - Revision Tibial Insert - Mod Rev/Dur Rev | 6637-0-928 6637-0-328 6637-0-631 6637-0-028 |
The industry guidance for storage duration of Ultra High Molecular Weight Polyethylene (UHMWPE) raw material used in the manufacture of soft implant bearing/articulating material of prosthetic joints is less than 5 years. Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties. Stryker has become aware that the devices listed in the product/lot number attachment PFA 3306872 section of the customer letter have been manufactured with UHMWPE raw material aged over 5 years.
Recall Start Date: December 5, 2023